CLI Reference

The pyKOSMOS++ command-line interface provides a complete spectroscopic reduction workflow through the kosmos-reduce command and its subcommands.

Main Command

kosmos-reduce

Process KOSMOS longslit spectroscopy observations from raw FITS files to wavelength-calibrated 1D spectra.

Synopsis:

kosmos-reduce [OPTIONS] --input-dir INPUT --output-dir OUTPUT

Required Arguments:

--input-dir PATH

Input directory containing FITS files organized in subdirectories:

  • biases/: Bias frames (≥3 required)

  • flats/: Flat field frames (≥3 required)

  • arcs/: Arc lamp frames (≥1 required)

  • science/: Science target frames

--output-dir PATH

Output directory for reduced data products. Will be created if it doesn’t exist.

Optional Arguments:

--config PATH

Path to custom YAML configuration file. If not provided, uses default KOSMOS parameters.

--mode {batch,interactive}

Processing mode:

  • batch: Automatic processing of all detected traces (default)

  • interactive: Visual trace selection with matplotlib GUI

--max-traces N

Maximum number of traces to extract per science frame (default: 5)

--verbose

Enable verbose logging output

--quiet

Suppress non-essential output

--log-file PATH

Write detailed logs to specified file

--validate-only

Validate calibrations and inputs without processing science frames

--overwrite

Overwrite existing output files without prompting

Examples:

Basic reduction with default settings:

kosmos-reduce --input-dir data/2024-01-15 --output-dir reduced/

Custom configuration with verbose output:

kosmos-reduce --input-dir data/faint_galaxy \\
              --output-dir reduced/ \\
              --config custom_config.yaml \\
              --verbose

Interactive trace selection:

kosmos-reduce --input-dir data/multi_slit \\
              --output-dir reduced/ \\
              --mode interactive

Validation only (check calibrations):

kosmos-reduce --input-dir data/ --output-dir tmp/ --validate-only

Subcommands

calibrate

Generate master bias and flat calibration frames only, without processing science data.

Synopsis:

kosmos-reduce calibrate [OPTIONS] --input-dir INPUT --output-dir OUTPUT

Arguments:

--input-dir PATH

Directory containing biases/ and flats/ subdirectories

--output-dir PATH

Directory for calibration outputs (calibrations/ subdirectory created)

--method {median,mean}

Combination method (default: median)

Example:

kosmos-reduce calibrate --input-dir data/ --output-dir calibrations/

wavelength

Perform wavelength calibration on arc lamp frames independently.

Synopsis:

kosmos-reduce wavelength [OPTIONS] --input-arc ARC --lamp-type LAMP

Required Arguments:

--input-arc PATH

Path to arc lamp FITS file

--lamp-type {HeNeAr,Ar,Kr,Ne}

Arc lamp type for linelist selection

Optional Arguments:

--output PATH

Output FITS table for wavelength solution (default: wavelength_solution.fits)

--poly-order N

Polynomial order (3-7, default: auto-select with BIC)

--rms-threshold FLOAT

Maximum acceptable RMS residual in Angstroms (default: 0.1)

--min-lines N

Minimum number of identified lines required (default: 20)

--plot

Generate diagnostic plot (wavelength_fit.pdf)

Example:

kosmos-reduce wavelength --input-arc arc_001.fits \\
                         --lamp-type HeNeAr \\
                         --poly-order 5 \\
                         --plot

combine

Combine multiple reduced 1D spectra onto a common wavelength grid.

Synopsis:

kosmos-reduce combine [OPTIONS] --input-spectra PATTERN --output OUTPUT

Required Arguments:

--input-spectra PATTERN

Glob pattern for input spectrum FITS files (e.g., "spectra_1d/*.fits")

--output PATH

Output combined spectrum FITS file

Optional Arguments:

--method {median,mean,weighted}

Combination method:

  • median: Robust to outliers

  • mean: Simple average

  • weighted: Weighted by inverse variance

--wavelength-grid START STOP STEP

Common wavelength grid in Angstroms (default: auto-determine from inputs)

--resolution FLOAT

Target spectral resolution in Angstroms (default: preserve native)

Example:

kosmos-reduce combine --input-spectra "reduced/spectra_1d/*.fits" \\
                     --output combined_spectrum.fits \\
                     --method weighted

Exit Codes

The pipeline uses standard exit codes to indicate processing status:

Code

Meaning

0

Success - all processing completed without errors

1

Missing or insufficient calibration frames

2

Invalid FITS files or configuration

3

Wavelength solution failed to converge

4

No traces detected in science frames

5

User canceled interactive mode

Example usage in scripts:

#!/bin/bash
kosmos-reduce --input-dir data/ --output-dir reduced/
EXIT_CODE=$?

if [ $EXIT_CODE -eq 0 ]; then
    echo "Reduction successful!"
elif [ $EXIT_CODE -eq 1 ]; then
    echo "Error: Missing calibration frames"
    exit 1
else
    echo "Pipeline error (code $EXIT_CODE)"
    exit 1
fi

Output Products

The pipeline creates a standardized output directory structure:

output_dir/
├── calibrations/
│   ├── master_bias.fits
│   ├── master_flat.fits
│   └── bad_pixel_mask.fits
├── reduced_2d/
│   ├── science_001_2d.fits
│   └── science_002_2d.fits
├── spectra_1d/
│   ├── science_001_trace1.fits
│   ├── science_001_trace2.fits
│   └── science_002_trace1.fits
├── wavelength_solutions/
│   └── wavelength_solution_arc001.fits
├── quality_reports/
│   ├── science_001_quality.yaml
│   └── summary_report.txt
├── diagnostic_plots/
│   ├── wavelength_solution.png
│   ├── science_001_2d.png
│   └── science_001_trace1_profile.png
└── logs/
    └── reduction_log.txt

See Output Products for detailed format specifications.

Configuration Files

Custom configuration files use YAML format with nested sections:

# Detector parameters
detector:
  gain: 1.4           # e-/ADU
  readnoise: 3.7      # e-
  saturate: 58982     # ADU

# Wavelength calibration
wavelength:
  max_order: 7
  initial_dispersion: 1.0
  rms_threshold: 0.1

# Trace detection
trace_detection:
  expected_fwhm: 4.0
  min_snr: 3.0
  min_separation: 20

# Extraction
extraction:
  method: optimal     # or 'boxcar'
  aperture_width: 10
  sky_buffer: 30

# Quality thresholds
quality:
  min_snr: 5.0
  max_wavelength_rms: 0.2

See Configuration Reference for complete parameter documentation.

Performance Tips

Speed up calibration creation:

  • Use --method mean instead of median for faster (but less robust) combination

  • Process only necessary calibration frames:

    # Only combine first 5 bias/flat frames
    ls biases/*.fits | head -5 | xargs -I {} cp {} subset_biases/
    

Parallel processing:

  • Process multiple nights independently in separate terminals

  • Use --max-traces to limit extraction overhead for multi-slit observations

Memory optimization:

  • Process large datasets in batches rather than all at once

  • Use --quiet mode to reduce I/O overhead from logging

Disk space:

  • Diagnostic plots can be large; use --no-plots (if implemented) for production runs

  • Remove intermediate reduced_2d/ files after successful 1D extraction

Troubleshooting

“No FITS files found”

  • Check directory structure matches expected layout

  • Ensure files have .fits extension (not .fit or .FITS)

  • Verify FITS headers contain IMAGETYP keyword

“Insufficient calibration frames”

  • Need ≥3 bias, ≥3 flat, ≥1 arc minimum

  • Check subdirectory names match exactly: biases/, flats/, arcs/

“Wavelength solution failed”

  • Arc lamp exposure may be too short (need SNR >10 for line detection)

  • Try --lamp-type matching actual lamp used

  • Use wavelength subcommand with --plot to diagnose

“No traces detected”

  • Lower min_snr in config for faint sources

  • Use --mode interactive to visually verify 2D spectrum

  • Check flat field quality (illumination pattern)

For more issues, see Troubleshooting.